{"id":7944,"date":"2026-10-13T08:00:00","date_gmt":"2026-10-13T06:00:00","guid":{"rendered":"https:\/\/chemspx.com\/?p=7944"},"modified":"2026-09-01T15:45:02","modified_gmt":"2026-09-01T13:45:02","slug":"what-is-the-difference-between-linear-and-step-gradient-elution","status":"publish","type":"wiki_post","link":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/","title":{"rendered":"What is the difference between linear and step gradient elution?"},"content":{"rendered":"<p>Linear gradient elution changes solvent composition continuously and smoothly over time, while step gradient elution shifts solvent strength in discrete, abrupt jumps. The core difference lies in how each approach moves compounds off the stationary phase: linear gradients offer fine control over selectivity, while step gradients prioritize speed and throughput. The sections below unpack each mode in detail, covering when to use one over the other and how gradient shape affects the quality of your separation.<\/p>\n<h2>How does each gradient mode actually move compounds through a column?<\/h2>\n<p>In gradient elution chromatography, solvent strength increases over time to push retained compounds off the stationary phase and toward the detector or collection vessel. A <strong>linear gradient<\/strong> does this by steadily increasing the proportion of the strong solvent at a constant rate, so compounds experience a smooth, predictable change in elution conditions throughout the run. A <strong>step gradient<\/strong> does the same job in stages, jumping from one fixed solvent composition to the next at defined time points.<\/p>\n<p>In practice, a linear gradient might move from 5% to 95% organic modifier over 20 minutes at a constant rate. A step gradient covering the same range might hold at 20%, then jump to 50%, then to 80%, with each plateau lasting several minutes. The result is that compounds elute in clusters corresponding to each solvent step rather than spreading evenly across the run.<\/p>\n<p>Both approaches exploit the same fundamental principle: as solvent strength increases, the affinity between analyte and stationary phase weakens, and compounds begin to migrate. The difference is in how precisely and continuously that transition is managed.<\/p>\n<h2>What are the advantages of linear gradient elution?<\/h2>\n<p>Linear gradient elution offers superior resolution, reproducibility, and method transferability compared to step gradients. Because solvent strength changes continuously, compounds are separated based on subtle differences in their interaction with the stationary phase, making it the preferred mode for analytical HPLC and for preparative separations where peak purity is critical.<\/p>\n<p>Key advantages include:<\/p>\n<ul>\n <li><strong>Better resolution between closely eluting compounds<\/strong> &#8211; the gradual change in elution strength gives the column more opportunity to differentiate similar analytes<\/li>\n <li><strong>Smoother, more symmetrical peaks<\/strong> &#8211; continuous solvent transitions reduce the peak distortion that abrupt steps can introduce<\/li>\n <li><strong>Easier method development and transfer<\/strong> &#8211; linear gradients are straightforward to optimize by adjusting slope, and they translate reliably between instruments and column formats<\/li>\n <li><strong>Predictable retention behavior<\/strong> &#8211; modeling and simulation tools work well with linear gradients, supporting systematic method optimization<\/li>\n <li><strong>Consistent baseline<\/strong> &#8211; gradual composition changes are less likely to cause sudden UV baseline shifts that interfere with detection<\/li>\n<\/ul>\n<p>For research laboratories in the Benelux working on complex mixtures, impurity profiling, or natural product isolation, linear gradient elution is typically the starting point for any new separation method.<\/p>\n<h2>When is step gradient elution the better choice?<\/h2>\n<p>Step gradient elution is the better choice when speed, solvent economy, or sample throughput matters more than the finest possible resolution. It excels in preparative and process-scale chromatography where compounds differ significantly in polarity or hydrophobicity, meaning high resolution between adjacent peaks is not required to achieve acceptable purity.<\/p>\n<p>Step gradients are particularly well suited to:<\/p>\n<ol>\n <li><strong>Class-based separations<\/strong> &#8211; when the target compound belongs to a chemically distinct group that elutes cleanly at a specific solvent composition, a well-placed step can wash it off rapidly without a lengthy ramp<\/li>\n <li><strong>Protein and peptide purification<\/strong> &#8211; in ion-exchange and affinity chromatography, step gradients are standard practice for washing impurities and then eluting the target in a concentrated fraction<\/li>\n <li><strong>High-throughput screening workflows<\/strong> &#8211; shorter run times and faster column re-equilibration make step gradients attractive when processing large numbers of samples<\/li>\n <li><strong>Solvent reduction strategies<\/strong> &#8211; because the column spends less time at intermediate compositions, overall solvent consumption can be lower<\/li>\n<\/ol>\n<p>The trade-off is that step gradients are less forgiving. Compounds that happen to elute at or near a step transition may co-elute or produce distorted peaks, requiring careful placement of each step to avoid losing resolution where it matters most.<\/p>\n<h2>How does gradient shape affect resolution and peak purity?<\/h2>\n<p>Gradient shape directly controls how long each compound spends in the column under conditions that promote separation. A shallower slope or smaller step size gives the stationary phase more time to differentiate compounds, improving resolution. A steeper slope or larger step compresses elution times, reducing resolution but shortening the run.<\/p>\n<p>In linear gradient elution, the slope is the primary lever. A shallow gradient over a long time window spreads peaks across the chromatogram, maximizing resolution but increasing run time and solvent use. A steep gradient compresses everything into a shorter window, which is faster but risks co-elution of compounds with similar retention factors.<\/p>\n<p>In step gradient elution, peak purity depends heavily on where each step falls relative to the compounds of interest. If a step change occurs while two compounds are still co-migrating through the column, both will elute together in the same fraction. This is why step gradient methods require more upfront knowledge of the sample&#8217;s composition. Scouting runs or analytical data from a linear gradient method are often used first to map where compounds elute, and that information then informs the placement of each step.<\/p>\n<p>Peak shape is also affected: abrupt solvent changes can cause a phenomenon called &#8220;gradient breakthrough,&#8221; where a sharp front of strong solvent pushes compounds off the column faster than expected, leading to fronting or tailing peaks.<\/p>\n<h2>Can you convert a step gradient method into a linear gradient?<\/h2>\n<p>Yes, converting a step gradient method into a linear gradient is straightforward in principle. The linear gradient is drawn to pass through or approximate the solvent compositions used in each step, creating a continuous ramp that covers the same overall range. In practice, this usually means identifying the start and end compositions of the step method and connecting them with a single linear ramp, or using a segmented linear gradient that more closely mirrors the shape of the original steps.<\/p>\n<p>The conversion is useful when moving from a preparative or process context into an analytical validation workflow, or when a step method produces unacceptable peak distortion that a linear approach would resolve. It is also a common strategy when scaling a method from flash chromatography to HPLC, where instrument precision makes linear gradients more reproducible.<\/p>\n<p>One important consideration: the two methods will not produce identical chromatograms. The linear version typically spreads peaks more evenly and may resolve compounds that co-eluted in the step method, but it will also extend run time. Adjusting the gradient slope after conversion is usually necessary to rebalance speed and resolution.<\/p>\n<h2>Which gradient type is standard in flash chromatography?<\/h2>\n<p>Step gradients are the traditional standard in <a href=\"https:\/\/chemspx.com\/fr\/contact\/\">flash chromatography<\/a>, largely because early flash systems used manual solvent switching and could not deliver a continuously changing composition. Modern automated flash chromatography systems support both step and linear gradients, and linear gradients are increasingly common for complex separations.<\/p>\n<p>In a typical flash purification workflow, a step gradient might move from a low-polarity mixture to progressively higher polarity in three to five steps, washing off the target compound in one of the middle fractions. This approach works well for routine compound isolation after synthesis, where the compound of interest is chemically distinct from the main impurities.<\/p>\n<p>Linear gradients in flash chromatography offer better resolution for samples with multiple closely related compounds, such as diastereomers or structurally similar natural products. The choice between the two often comes down to the complexity of the sample: simple, well-characterized mixtures suit step gradients, while complex or unknown samples benefit from the finer control of a linear solvent gradient. Many modern preparative chromatography systems in Benelux research labs allow users to combine both approaches within a single run, using a step to load and wash, then switching to a linear ramp for the elution window where resolution matters most.<\/p>\n<h2>How ChemSPX supports your gradient elution workflow<\/h2>\n<p>Choosing the right gradient strategy is only part of the challenge. Getting consistent, reproducible results also depends on having the right instrumentation and the expertise to configure it correctly. ChemSPX helps research and analytical laboratories across the Benelux select, install, and optimize preparative and flash chromatography systems for both linear and step gradient applications.<\/p>\n<ul>\n <li>Application-level advice on gradient method development for your specific sample type<\/li>\n <li>Instrument selection and configuration for batch and preparative chromatography workflows<\/li>\n <li>On-site installation and qualification (IQ\/OQ) in line with GMP and GLP requirements<\/li>\n <li>Ongoing service and calibration support to keep your system performing to specification<\/li>\n <li>Access to a specialist team of Sales Engineers and Product Specialists based in the Netherlands and Belgium<\/li>\n<\/ul>\n<p>If you are evaluating gradient elution systems or looking to improve an existing separation workflow, <a href=\"https:\/\/chemspx.com\/fr\/contact\/\">contact ChemSPX<\/a> to speak with a specialist.<\/p>\n<div class=\"wp-block-seoaic-faq-block\">\n    <h2 class=\"seoaic-faq-section-title\">Foire aux questions<\/h2>\n            <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                How do I know if my current step gradient method is causing co-elution problems?            <\/h3>\n            <p class=\"seoaic-answer\">\n                The most reliable way to diagnose co-elution in a step gradient method is to run the same sample using a shallow linear gradient across the full solvent range and compare the resulting chromatogram. If the linear run reveals additional peaks or shoulder peaks that were hidden in the step method, co-elution is occurring. Fraction analysis by LC-MS or NMR can also confirm whether collected fractions contain more than one compound, even when the chromatogram appears clean.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What is the best way to get started with linear gradient method development if I have no prior data on my sample?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Start with a broad scouting gradient \u2014 typically running from 5% to 95% strong solvent over 20\u201330 minutes \u2014 to map where your compounds elute across the full retention range. Once you know the elution window, you can narrow the gradient to focus on that region, shallowing the slope to improve resolution between closely spaced peaks. Most method development guidelines recommend this scouting approach before investing time in fine optimization.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                Can gradient elution damage or prematurely wear out my column?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Gradient elution itself does not inherently damage columns, but abrupt step changes can cause mechanical stress on the stationary phase bed if the solvent viscosity changes significantly between steps, particularly in normal-phase systems. Ensuring adequate re-equilibration time between runs is critical \u2014 insufficient re-equilibration is one of the most common causes of retention time drift and reduced column lifetime in gradient methods. Always follow the column manufacturer&#8217;s recommended equilibration volumes for the solvent system you are using.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                Why does my baseline shift dramatically during a gradient run, and how can I fix it?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Baseline shifts during gradient elution are most commonly caused by differences in UV absorbance between the weak and strong solvents, or by trace impurities in the mobile phase that accumulate and elute as the solvent composition changes. Using HPLC-grade solvents and filtering them through 0.2 \u00b5m membranes reduces impurity-driven baseline noise. For UV-absorbing solvents like THF or acetone, switching to a low-UV-absorbing alternative such as acetonitrile, or using a reference wavelength, can significantly flatten the baseline.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                How long should I re-equilibrate my column between gradient runs to ensure reproducible retention times?            <\/h3>\n            <p class=\"seoaic-answer\">\n                A general rule of thumb is to re-equilibrate with at least 10 column volumes of the starting mobile phase composition before the next injection, though some stationary phases \u2014 particularly those with ion-pairing reagents or high aqueous content \u2014 may require 20 or more column volumes. Insufficient re-equilibration is one of the most frequent sources of retention time variability in gradient methods. If you are seeing drift across a sequence of injections, extending re-equilibration time is the first variable to adjust before investigating other causes.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                Is it possible to use a combination of step and linear gradient within a single chromatographic run?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Yes, and this hybrid approach is increasingly common in modern preparative and flash chromatography workflows. A typical strategy uses an initial step to rapidly wash off highly polar or non-retained impurities, followed by a linear gradient ramp through the elution window where target compounds and closely related impurities need to be resolved, and then a final steep step or wash to strip strongly retained material from the column. Most modern automated chromatography systems support multi-segment gradient programming that makes this straightforward to implement.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What common mistakes should I avoid when transferring a gradient method from one instrument to another?            <\/h3>\n            <p class=\"seoaic-answer\">\n                The most critical factors to account for during method transfer are differences in system dwell volume (the volume between the gradient mixer and the column head) and flow rate accuracy between instruments. A higher dwell volume on the receiving instrument will delay the start of the gradient relative to the injection, shifting retention times and potentially altering selectivity. To correct for this, measure the dwell volume of both systems and adjust the gradient start time or add a pre-gradient isocratic hold on the new instrument to compensate for the difference.            <\/p>\n        <\/div>\n        <\/div>","protected":false},"template":"","class_list":["post-7944","wiki_post","type-wiki_post","status-publish","has-post-thumbnail","hentry"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v28.6 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>What is the difference between linear and step gradient elution? | ChemSPX<\/title>\n<meta name=\"description\" content=\"Linear and step gradient elution differ in control, speed, and resolution. Learn which method suits your separation and how gradient shape affects peak purity.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/\" \/>\n<meta property=\"og:locale\" content=\"fr_FR\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"What is the difference between linear and step gradient elution? | ChemSPX\" \/>\n<meta property=\"og:description\" content=\"Linear and step gradient elution differ in control, speed, and resolution. Learn which method suits your separation and how gradient shape affects peak purity.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/\" \/>\n<meta property=\"og:site_name\" content=\"ChemSPX\" \/>\n<meta property=\"og:image\" content=\"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-separated-compounds-laboratory.webp\" \/>\n\t<meta property=\"og:image:width\" content=\"1024\" \/>\n\t<meta property=\"og:image:height\" content=\"768\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/webp\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Dur\u00e9e de lecture estim\u00e9e\" \/>\n\t<meta name=\"twitter:data1\" content=\"11 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/\",\"url\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/\",\"name\":\"What is the difference between linear and step gradient elution? | ChemSPX\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/09\\\/chromatography-column-separated-compounds-laboratory.webp\",\"datePublished\":\"2026-10-13T06:00:00+00:00\",\"description\":\"Linear and step gradient elution differ in control, speed, and resolution. Learn which method suits your separation and how gradient shape affects peak purity.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/#breadcrumb\"},\"inLanguage\":\"fr-FR\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"fr-FR\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/#primaryimage\",\"url\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/09\\\/chromatography-column-separated-compounds-laboratory.webp\",\"contentUrl\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/09\\\/chromatography-column-separated-compounds-laboratory.webp\",\"width\":1024,\"height\":768,\"caption\":\"Glass chromatography column separating compounds into vivid violet, amber, and teal bands on a steel lab bench with solvent gradient tubing.\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/wiki\\\/what-is-the-difference-between-linear-and-step-gradient-elution\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/chemspx.com\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Wiki post\",\"item\":\"https:\\\/\\\/chemspx.com\\\/Wiki\\\/\"},{\"@type\":\"ListItem\",\"position\":3,\"name\":\"What is the difference between linear and step gradient elution?\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/#website\",\"url\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/\",\"name\":\"ChemSPX\",\"description\":\"ChemSPX\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/chemspx.com\\\/fr\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"fr-FR\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"What is the difference between linear and step gradient elution? | ChemSPX","description":"Linear and step gradient elution differ in control, speed, and resolution. Learn which method suits your separation and how gradient shape affects peak purity.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/","og_locale":"fr_FR","og_type":"article","og_title":"What is the difference between linear and step gradient elution? | ChemSPX","og_description":"Linear and step gradient elution differ in control, speed, and resolution. Learn which method suits your separation and how gradient shape affects peak purity.","og_url":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/","og_site_name":"ChemSPX","og_image":[{"width":1024,"height":768,"url":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-separated-compounds-laboratory.webp","type":"image\/webp"}],"twitter_card":"summary_large_image","twitter_misc":{"Dur\u00e9e de lecture estim\u00e9e":"11 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"WebPage","@id":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/","url":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/","name":"What is the difference between linear and step gradient elution? | ChemSPX","isPartOf":{"@id":"https:\/\/chemspx.com\/fr\/#website"},"primaryImageOfPage":{"@id":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/#primaryimage"},"image":{"@id":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/#primaryimage"},"thumbnailUrl":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-separated-compounds-laboratory.webp","datePublished":"2026-10-13T06:00:00+00:00","description":"Linear and step gradient elution differ in control, speed, and resolution. Learn which method suits your separation and how gradient shape affects peak purity.","breadcrumb":{"@id":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/#breadcrumb"},"inLanguage":"fr-FR","potentialAction":[{"@type":"ReadAction","target":["https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/"]}]},{"@type":"ImageObject","inLanguage":"fr-FR","@id":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/#primaryimage","url":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-separated-compounds-laboratory.webp","contentUrl":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-separated-compounds-laboratory.webp","width":1024,"height":768,"caption":"Glass chromatography column separating compounds into vivid violet, amber, and teal bands on a steel lab bench with solvent gradient tubing."},{"@type":"BreadcrumbList","@id":"https:\/\/chemspx.com\/fr\/wiki\/what-is-the-difference-between-linear-and-step-gradient-elution\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/chemspx.com\/"},{"@type":"ListItem","position":2,"name":"Wiki post","item":"https:\/\/chemspx.com\/Wiki\/"},{"@type":"ListItem","position":3,"name":"What is the difference between linear and step gradient elution?"}]},{"@type":"WebSite","@id":"https:\/\/chemspx.com\/fr\/#website","url":"https:\/\/chemspx.com\/fr\/","name":"ChemSPX","description":"ChemSPX","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/chemspx.com\/fr\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"fr-FR"}]}},"_links":{"self":[{"href":"https:\/\/chemspx.com\/fr\/wp-json\/wp\/v2\/wiki_post\/7944","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/chemspx.com\/fr\/wp-json\/wp\/v2\/wiki_post"}],"about":[{"href":"https:\/\/chemspx.com\/fr\/wp-json\/wp\/v2\/types\/wiki_post"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/chemspx.com\/fr\/wp-json\/wp\/v2\/media\/7977"}],"wp:attachment":[{"href":"https:\/\/chemspx.com\/fr\/wp-json\/wp\/v2\/media?parent=7944"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}