{"id":7951,"date":"2026-10-09T08:00:00","date_gmt":"2026-10-09T06:00:00","guid":{"rendered":"https:\/\/chemspx.com\/?p=7951"},"modified":"2026-09-01T15:45:01","modified_gmt":"2026-09-01T13:45:01","slug":"how-do-you-assess-column-lifetime-and-decide-when-to-repack","status":"publish","type":"wiki_post","link":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/","title":{"rendered":"How do you assess column lifetime and decide when to repack?"},"content":{"rendered":"<p>You should assess column lifetime by tracking key performance indicators over time and decide to repack when those indicators fall outside acceptable limits. The most reliable signals are a significant increase in pressure drop, loss of separation efficiency, peak tailing, and reduced resolution between target compounds. These changes indicate that the packed bed has degraded beyond what regeneration can recover. The sections below address each diagnostic step and decision point in detail.<\/p>\n<h2>What are the signs that a chromatography column is failing?<\/h2>\n<p>A failing chromatography column typically shows one or more of four key symptoms: rising backpressure, deteriorating peak shape, reduced resolution, and declining yield of the target compound. These signs can appear gradually or accelerate suddenly depending on the cause of degradation.<\/p>\n<p>In preparative chromatography and flash chromatography, the most visible early warning is peak broadening. When compounds that previously eluted as sharp, symmetrical peaks begin to tail or merge with adjacent peaks, the stationary phase is no longer performing uniformly. Void formation at the top of the column is a common physical cause, creating channels through which mobile phase bypasses the packed bed.<\/p>\n<p>Rising column pressure drop is another reliable indicator. If the pressure required to maintain a set flow rate increases noticeably from run to run, fines accumulation, bed compression, or biological fouling may be responsible. Monitoring pressure trends over time is one of the most practical ways to detect early-stage column degradation before separation quality is compromised.<\/p>\n<h2>How do you measure column performance over time?<\/h2>\n<p>Column performance is measured by calculating theoretical plate count (N), peak asymmetry factor, and resolution (Rs) using a standard test compound under defined conditions. Running these tests at regular intervals and comparing results against an initial baseline gives a clear picture of how the column is aging.<\/p>\n<p>A structured performance log is the most practical tool for this. After each defined number of runs or volume of mobile phase processed, inject the test compound and record:<\/p>\n<ul>\n <li>Theoretical plate count (N) &#8211; a measure of separation efficiency<\/li>\n <li>Peak asymmetry or tailing factor &#8211; indicates bed homogeneity<\/li>\n <li>Resolution between two reference compounds &#8211; reflects overall selectivity<\/li>\n <li>Operating pressure at a fixed flow rate &#8211; flags physical bed changes<\/li>\n <li>Column back-pressure trend over time &#8211; an early warning of fouling or compression<\/li>\n<\/ul>\n<p>When any of these values drifts more than 20 to 30 percent from the original baseline, the column warrants closer investigation. In regulated environments, the acceptance criteria are typically defined in advance as part of the column qualification protocol.<\/p>\n<h2>What factors shorten the lifetime of a packed column?<\/h2>\n<p>The main factors that shorten column lifetime are sample matrix contamination, incompatible mobile phase conditions, mechanical stress from pressure cycling, and inadequate flushing between runs. Each of these degrades the stationary phase or the physical integrity of the packed bed in a different way.<\/p>\n<p>Strongly retained matrix components, such as lipids, proteins, or particulates, accumulate at the column inlet and gradually block the bed. Mobile phases outside the pH range recommended for the stationary phase chemically degrade the bonded phase or dissolve the silica support. Repeated pressure spikes, especially during column switching or pump startup, cause mechanical compression of the bed over time.<\/p>\n<p>In preparative and flash chromatography applications, sample loading is a particularly significant factor. Overloading the column beyond its designed capacity accelerates stationary phase wear and increases the risk of irreversible fouling. Consistent use of inline filters and guard columns reduces the rate of contamination reaching the main bed.<\/p>\n<h2>Can a column be regenerated instead of repacked?<\/h2>\n<p>Yes, in many cases a column can be regenerated rather than repacked. Regeneration involves flushing the column with a sequence of solvents chosen to dissolve and remove the accumulated contaminants without replacing the stationary phase. Whether regeneration is sufficient depends on the nature of the fouling and the condition of the packed bed itself.<\/p>\n<p>If performance loss is caused by reversible fouling, such as retained hydrophobic compounds, salts, or lipids, a targeted wash protocol can restore much of the original efficiency. Common approaches include flushing with strong organic solvents, aqueous buffers at adjusted pH, or dedicated stripping solutions compatible with the stationary phase.<\/p>\n<p>However, if the bed has physically collapsed, developed voids, or if the stationary phase has chemically degraded, regeneration will not restore performance. In those cases, repacking is the appropriate next step. Regeneration is most effective as a preventive measure applied routinely, rather than as a rescue strategy after performance has severely declined.<\/p>\n<h2>How do you decide when repacking is the right choice?<\/h2>\n<p>Repacking is the right choice when column performance cannot be restored by regeneration, when physical bed damage is confirmed, or when the cost of continued poor separation outweighs the cost of repacking. The decision should be based on measured data, not on run count alone.<\/p>\n<p>A structured decision process typically follows these steps:<\/p>\n<ol>\n <li>Run a performance test and compare results against the original baseline.<\/li>\n <li>Attempt regeneration if fouling is the suspected cause and the bed appears physically intact.<\/li>\n <li>Re-test after regeneration to determine whether performance has recovered.<\/li>\n <li>Inspect the column visually or by pressure mapping if performance remains poor.<\/li>\n <li>Repack if the bed shows voids, channeling, or irreversible phase degradation.<\/li>\n <li>Document the decision and the performance data for future reference.<\/li>\n<\/ol>\n<p>In environments where <a href=\"https:\/\/chemspx.com\/nl\/contact\/\">column repacking support<\/a> is available from a specialist, it is worth consulting before committing to a full repack. An experienced application specialist can sometimes identify a targeted fix that avoids the time and cost of a full repacking cycle.<\/p>\n<h2>What does the repacking process involve?<\/h2>\n<p>Repacking a chromatography column involves emptying the column housing, cleaning it thoroughly, preparing a fresh slurry of stationary phase, and packing it under controlled pressure to achieve a homogeneous, void-free bed. The process requires care and the right equipment to produce a column that performs comparably to a factory-packed unit.<\/p>\n<p>The stationary phase slurry is prepared by suspending the packing material in an appropriate solvent at a defined concentration. The slurry is then pushed into the column at a controlled flow rate and pressure using a packing pump or reservoir. After packing, the bed is consolidated by pumping mobile phase through at the intended operating pressure, and a performance test is run to confirm the column meets its specification before returning it to service.<\/p>\n<p>For preparative and flash chromatography columns, repacking in-house is common practice and can be cost-effective when the hardware is in good condition. The stationary phase itself is typically the main consumable cost. Proper technique and equipment calibration are essential to achieving reproducible results.<\/p>\n<h2>How can column lifetime be extended from the start?<\/h2>\n<p>Column lifetime is extended by using inline filters, applying consistent flushing protocols, avoiding pressure spikes, and storing columns correctly when not in use. These practices reduce the rate at which the stationary phase and bed structure degrade, making each column investment last significantly longer.<\/p>\n<p>Good habits established at the start of a column&#8217;s working life have a compounding effect. Filtering samples before injection prevents particulate contamination from reaching the bed. Equilibrating the column gradually rather than applying full flow rate immediately reduces mechanical stress. Flushing with appropriate solvents at the end of each session removes retained compounds before they become difficult to displace.<\/p>\n<p>Storage conditions also matter. Columns left in aqueous mobile phases are vulnerable to microbial growth; storing them in an appropriate organic solvent prevents this. Keeping columns sealed and away from temperature extremes preserves both the stationary phase chemistry and the physical integrity of the bed. Consistent, documented maintenance routines are the most reliable way to maximize column lifetime in any preparative chromatography workflow.<\/p>\n<h2>How ChemSPX supports column lifetime management<\/h2>\n<p>ChemSPX helps laboratories across the Benelux get the most from their preparative and flash chromatography columns, from initial setup through long-term maintenance. The team of Sales Engineers and Product Specialists works directly with lab professionals to advise on column selection, performance monitoring, and maintenance strategies suited to their specific workflows and sample types.<\/p>\n<ul>\n <li>Advice on column qualification, performance testing, and baseline documentation<\/li>\n <li>Guidance on regeneration protocols and when repacking is the right decision<\/li>\n <li>Support for IQ\/OQ installations and compliance-aligned maintenance in regulated environments<\/li>\n <li>Application-level assistance for chromatography workflows in pharmaceutical, chemical, and research settings<\/li>\n<\/ul>\n<p>If you are seeing signs of column degradation or want to set up a structured maintenance program for your chromatography systems, <a href=\"https:\/\/chemspx.com\/nl\/contact\/\">contact ChemSPX<\/a> to speak with a specialist.<\/p>\n<div class=\"wp-block-seoaic-faq-block\">\n    <h2 class=\"seoaic-faq-section-title\">Veelgestelde vragen<\/h2>\n            <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                How often should I run a formal performance test on my preparative chromatography column?            <\/h3>\n            <p class=\"seoaic-answer\">\n                The ideal testing frequency depends on your throughput and application, but a practical starting point is to run a full performance test every 20\u201350 injections, or after every significant change in sample type or mobile phase composition. In high-throughput or regulated environments, testing intervals are typically defined in the column qualification protocol and tied to cumulative volume processed rather than run count alone. The key is consistency \u2014 testing at irregular intervals makes trend data unreliable and reduces your ability to catch early-stage degradation before it affects product quality.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What is the difference between a guard column and an inline filter, and do I need both?            <\/h3>\n            <p class=\"seoaic-answer\">\n                A guard column is a short, replaceable column packed with the same stationary phase as your main column, designed to capture strongly retained matrix components before they reach the main bed. An inline filter, by contrast, is a physical barrier that removes particulates from the mobile phase or sample stream. Both serve different protective functions, and using them together provides the most comprehensive protection \u2014 the filter prevents particulate fouling while the guard column absorbs chemical contaminants. For preparative and flash chromatography applications where sample matrices are complex or variable, using both is strongly recommended and significantly extends main column lifetime.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                Can I reuse the stationary phase after repacking, or does it need to be replaced with fresh material?            <\/h3>\n            <p class=\"seoaic-answer\">\n                In some cases, stationary phase can be recovered and reused, but this depends on the extent of chemical degradation and the nature of any retained contaminants. If the phase has been exposed to conditions outside its recommended pH range, or if it shows visible discoloration or clumping, it should be replaced rather than reused. When reuse is considered, the recovered material should be thoroughly washed, dried, and characterized before repacking, and a post-pack performance test should be run to confirm it meets the original specification. For regulated applications, reusing stationary phase typically requires documented justification and additional quality controls.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What should I do if column pressure drops suddenly rather than rising gradually?            <\/h3>\n            <p class=\"seoaic-answer\">\n                A sudden drop in backpressure \u2014 rather than the more common gradual increase \u2014 often signals a physical failure such as a cracked bed, a loose end fitting, or a fractured frit that is no longer providing adequate resistance. This is distinct from fouling-related pressure increases and usually cannot be resolved by regeneration. The column should be taken offline immediately, inspected visually, and pressure-tested to identify the source. If the bed has collapsed or the hardware is compromised, repacking or column replacement is the appropriate response rather than attempting further regeneration.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                How do I know if peak tailing is caused by the column or by something else in my system?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Before attributing peak tailing to column degradation, it is worth ruling out extra-column contributions such as dead volume in connecting tubing, detector cell volume, or poorly optimized injection conditions. A quick diagnostic is to test the same column on a different system, or to inject a small-molecule standard under isocratic conditions with minimal tubing \u2014 if tailing disappears or improves significantly, the issue is likely system-related rather than column-related. If tailing persists across systems and worsens over time relative to your baseline, the packed bed is the more probable cause, and a regeneration trial followed by re-testing is the recommended next step.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What is the correct way to store a preparative chromatography column that will not be used for several weeks?            <\/h3>\n            <p class=\"seoaic-answer\">\n                For storage periods of more than a few days, columns should be flushed thoroughly to remove any sample residues or aqueous buffers, then stored filled with an appropriate organic solvent \u2014 typically the one recommended by the stationary phase manufacturer, often acetonitrile or methanol for reversed-phase media. Both end fittings should be capped tightly to prevent solvent evaporation and bed drying, which can cause shrinkage and void formation. Columns should be stored upright at room temperature or as specified, away from direct sunlight and temperature extremes, and a brief equilibration run should be performed before returning the column to service.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                At what point does it make more economic sense to replace the column hardware rather than repacking it?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Repacking is cost-effective as long as the column hardware \u2014 the column body, end fittings, frits, and seals \u2014 is in good mechanical condition and free from corrosion, cracking, or deformation. If the hardware requires replacement of multiple components, shows signs of metal fatigue, or has been exposed to chemically incompatible solvents that have degraded seals or coatings, the cumulative cost of refurbishment can approach or exceed the cost of new hardware. A practical rule of thumb is to inspect hardware condition at every repacking cycle and compare the cost of necessary repairs against current replacement pricing \u2014 in many preparative setups, column bodies last through multiple repacking cycles before hardware replacement becomes the more economical choice.            <\/p>\n        <\/div>\n        <\/div>","protected":false},"template":"","class_list":["post-7951","wiki_post","type-wiki_post","status-publish","has-post-thumbnail","hentry"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v28.6 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>How do you assess column lifetime and decide when to repack? | ChemSPX<\/title>\n<meta name=\"description\" content=\"Track column performance, spot early failure signs, and know when regeneration is not enough. Learn how to decide when repacking is the right call.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/\" \/>\n<meta property=\"og:locale\" content=\"nl_NL\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"How do you assess column lifetime and decide when to repack? | ChemSPX\" \/>\n<meta property=\"og:description\" content=\"Track column performance, spot early failure signs, and know when regeneration is not enough. Learn how to decide when repacking is the right call.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/\" \/>\n<meta property=\"og:site_name\" content=\"ChemSPX\" \/>\n<meta property=\"og:image\" content=\"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-worn-resin-lab-inspection.webp\" \/>\n\t<meta property=\"og:image:width\" content=\"1024\" \/>\n\t<meta property=\"og:image:height\" content=\"768\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/webp\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Geschatte leestijd\" \/>\n\t<meta name=\"twitter:data1\" content=\"11 minuten\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/\",\"url\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/\",\"name\":\"How do you assess column lifetime and decide when to repack? | ChemSPX\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/09\\\/chromatography-column-worn-resin-lab-inspection.webp\",\"datePublished\":\"2026-10-09T06:00:00+00:00\",\"description\":\"Track column performance, spot early failure signs, and know when regeneration is not enough. Learn how to decide when repacking is the right call.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/#breadcrumb\"},\"inLanguage\":\"nl-NL\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"nl-NL\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/#primaryimage\",\"url\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/09\\\/chromatography-column-worn-resin-lab-inspection.webp\",\"contentUrl\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/09\\\/chromatography-column-worn-resin-lab-inspection.webp\",\"width\":1024,\"height\":768,\"caption\":\"Gloved hand examining a worn chromatography column with darkened, unevenly packed resin on a laboratory bench.\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/chemspx.com\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Wiki post\",\"item\":\"https:\\\/\\\/chemspx.com\\\/Wiki\\\/\"},{\"@type\":\"ListItem\",\"position\":3,\"name\":\"How do you assess column lifetime and decide when to repack?\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/#website\",\"url\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/\",\"name\":\"ChemSPX\",\"description\":\"ChemSPX\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"nl-NL\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"How do you assess column lifetime and decide when to repack? | ChemSPX","description":"Track column performance, spot early failure signs, and know when regeneration is not enough. Learn how to decide when repacking is the right call.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/","og_locale":"nl_NL","og_type":"article","og_title":"How do you assess column lifetime and decide when to repack? | ChemSPX","og_description":"Track column performance, spot early failure signs, and know when regeneration is not enough. Learn how to decide when repacking is the right call.","og_url":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/","og_site_name":"ChemSPX","og_image":[{"width":1024,"height":768,"url":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-worn-resin-lab-inspection.webp","type":"image\/webp"}],"twitter_card":"summary_large_image","twitter_misc":{"Geschatte leestijd":"11 minuten"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"WebPage","@id":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/","url":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/","name":"How do you assess column lifetime and decide when to repack? | ChemSPX","isPartOf":{"@id":"https:\/\/chemspx.com\/nl\/#website"},"primaryImageOfPage":{"@id":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/#primaryimage"},"image":{"@id":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/#primaryimage"},"thumbnailUrl":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-worn-resin-lab-inspection.webp","datePublished":"2026-10-09T06:00:00+00:00","description":"Track column performance, spot early failure signs, and know when regeneration is not enough. Learn how to decide when repacking is the right call.","breadcrumb":{"@id":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/#breadcrumb"},"inLanguage":"nl-NL","potentialAction":[{"@type":"ReadAction","target":["https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/"]}]},{"@type":"ImageObject","inLanguage":"nl-NL","@id":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/#primaryimage","url":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-worn-resin-lab-inspection.webp","contentUrl":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/09\/chromatography-column-worn-resin-lab-inspection.webp","width":1024,"height":768,"caption":"Gloved hand examining a worn chromatography column with darkened, unevenly packed resin on a laboratory bench."},{"@type":"BreadcrumbList","@id":"https:\/\/chemspx.com\/nl\/wiki\/how-do-you-assess-column-lifetime-and-decide-when-to-repack\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/chemspx.com\/"},{"@type":"ListItem","position":2,"name":"Wiki post","item":"https:\/\/chemspx.com\/Wiki\/"},{"@type":"ListItem","position":3,"name":"How do you assess column lifetime and decide when to repack?"}]},{"@type":"WebSite","@id":"https:\/\/chemspx.com\/nl\/#website","url":"https:\/\/chemspx.com\/nl\/","name":"ChemSPX","description":"ChemSPX","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/chemspx.com\/nl\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"nl-NL"}]}},"_links":{"self":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/wiki_post\/7951","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/wiki_post"}],"about":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/types\/wiki_post"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/media\/7984"}],"wp:attachment":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/media?parent=7951"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}