{"id":8111,"date":"2026-10-24T08:00:00","date_gmt":"2026-10-24T06:00:00","guid":{"rendered":"https:\/\/chemspx.com\/?p=8111"},"modified":"2026-10-13T10:28:21","modified_gmt":"2026-10-13T08:28:21","slug":"what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations","status":"publish","type":"wiki_post","link":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/","title":{"rendered":"What is the difference between IVRT and IVPT for topical formulations?"},"content":{"rendered":"<p>IVRT (in vitro release testing) and IVPT (in vitro permeation testing) are two distinct analytical methods used to evaluate topical and semisolid formulations, but they measure fundamentally different things. IVRT measures how a drug releases from its formulation matrix, while IVPT measures how much of that drug actually crosses a biological membrane into underlying tissue. The choice between them depends on the regulatory purpose, the formulation type, and the stage of development. The sections below unpack each question in detail.<\/p>\n<h2>What does each test actually measure?<\/h2>\n<p>IVRT measures the rate and extent of release of an active pharmaceutical ingredient from a semisolid formulation such as a cream, gel, or ointment. It evaluates the formulation itself, not the biological response. IVPT, by contrast, measures the permeation of a drug through a membrane intended to represent skin, quantifying how much drug crosses into a receptor fluid over time.<\/p>\n<p>This distinction matters because the two tests answer different scientific questions. IVRT asks: is the drug being made available from the product? IVPT asks: once available, how much reaches the target tissue? In pharmaceutical research, IVRT is often used to assess formulation consistency and detect changes in manufacturing, while IVPT provides insight into transdermal drug delivery potential and skin absorption behavior.<\/p>\n<p>Both tests are conducted under controlled temperature conditions and use a diffusion cell setup, but the endpoint and interpretation differ significantly. A formulation can release drug efficiently in IVRT while still showing limited permeation in IVPT, depending on the physicochemical properties of the active ingredient and the nature of the membrane used.<\/p>\n<h2>What membranes are used in IVRT and IVPT?<\/h2>\n<p>IVRT typically uses a synthetic, inert membrane such as a cellulose acetate or polysulfone membrane. These membranes do not act as a barrier themselves; their role is simply to support the formulation and prevent it from mixing with the receptor fluid below. The membrane in IVRT is not meant to simulate skin.<\/p>\n<p>IVPT uses membranes that are intended to represent human skin. The most commonly used options include:<\/p>\n<ul>\n <li><strong>Excised human skin<\/strong>: considered the gold standard, sourced from surgical or cadaver tissue<\/li>\n <li><strong>Excised animal skin<\/strong>: pig ear skin is widely used due to its structural similarity to human skin<\/li>\n <li><strong>Reconstructed human epidermis models<\/strong>: engineered tissue constructs that offer reproducibility and eliminate donor variability<\/li>\n<\/ul>\n<p>The choice of membrane in IVPT directly influences the permeation data and its relevance to in vivo conditions. Human skin is preferred for regulatory submissions, but animal and reconstructed models are frequently used in early-stage research and formulation screening due to their greater availability and consistency.<\/p>\n<h2>When is IVRT required versus IVPT?<\/h2>\n<p>IVRT is generally required when demonstrating that a generic topical formulation is equivalent to a reference listed drug, particularly for locally acting products where systemic absorption is not the primary concern. Regulatory guidance in several jurisdictions specifies IVRT as the in vitro component of a bioequivalence strategy for semisolid formulations.<\/p>\n<p>IVPT is typically required or preferred when the goal is to understand transdermal absorption, support safety assessments, or characterize how a drug penetrates through skin layers into deeper tissue or systemic circulation. It is also used in early-stage research to rank and screen formulation candidates before moving into more resource-intensive in vivo studies.<\/p>\n<p>In practice, the two tests often complement each other. IVRT establishes that the formulation consistently delivers the drug to the membrane surface, while IVPT adds information about what happens after that release occurs. Research laboratories working on novel topical or transdermal formulations frequently run both tests at different stages of development.<\/p>\n<h2>How does the Franz diffusion cell work in both tests?<\/h2>\n<p>The Franz diffusion cell is the standard apparatus used in both IVRT and IVPT. It consists of two chambers separated by a membrane: a donor compartment on top, where the formulation is applied, and a receptor compartment below, filled with a fluid that mimics physiological conditions. Drug that releases from the formulation and crosses the membrane accumulates in the receptor fluid and is sampled at defined time intervals.<\/p>\n<p>The cell is maintained at a controlled temperature, typically 32\u00b0C for skin surface simulation or 37\u00b0C for systemic models, using a water jacket or heating block. Stirring in the receptor compartment ensures that drug does not accumulate at the membrane surface, maintaining sink conditions throughout the experiment.<\/p>\n<p>The key operational difference between IVRT and IVPT lies in what sits between the donor and receptor compartments. In IVRT, an inert synthetic membrane holds the formulation in place. In IVPT, a skin membrane replaces that inert layer, introducing a genuine biological barrier. The data collected, drug concentration in the receptor fluid over time, is analyzed in both cases, but the interpretation reflects the nature of the membrane used.<\/p>\n<h2>Can IVRT and IVPT results be compared or correlated?<\/h2>\n<p>IVRT and IVPT results cannot be directly compared because they measure different phenomena, but correlations between the two can be established under specific conditions. If a formulation change increases drug release in IVRT, it may also increase permeation in IVPT, but this relationship is not automatic and depends on the rate-limiting step in the overall transport process.<\/p>\n<p>When the formulation itself is the rate-limiting step, meaning the drug struggles to leave the matrix rather than to cross the skin, IVRT and IVPT results tend to track each other more closely. When the skin membrane is the dominant barrier, IVRT results may not predict IVPT outcomes reliably.<\/p>\n<p>Establishing an in vitro-in vitro correlation between IVRT and IVPT data can be valuable in formulation research. It allows scientists to use the faster and simpler IVRT method as a screening tool and reserve IVPT for candidates that pass the release threshold. This staged approach is common in Benelux pharmaceutical and cosmetic research laboratories where throughput and resource efficiency are important considerations.<\/p>\n<h2>Which instruments are used for IVRT and IVPT in the lab?<\/h2>\n<p>The core instrument for both IVRT and IVPT is the Franz diffusion cell system, which may be used as individual static cells or as an automated multi-cell apparatus. Automated systems allow simultaneous sampling from multiple cells, improving throughput and reducing manual handling errors.<\/p>\n<p>A complete laboratory setup for IVRT and IVPT typically includes:<\/p>\n<ol>\n <li><strong>Franz diffusion cells<\/strong>: static or flow-through, glass or composite construction<\/li>\n <li><strong>Temperature-controlled water bath or heating block<\/strong>: to maintain precise thermal conditions throughout the experiment<\/li>\n <li><strong>Automated sampling system<\/strong>: for timed collection of receptor fluid fractions without disrupting the experiment<\/li>\n <li><strong>Analytical instrument<\/strong>: typically an HPLC or UV-Vis spectrophotometer to quantify drug concentration in collected samples<\/li>\n <li><strong>Stirring mechanism<\/strong>: magnetic stirrers or paddle systems integrated into the cell block to maintain sink conditions<\/li>\n<\/ol>\n<p>Laboratories running GMP-compliant or GLP-compliant studies require instruments that support calibration and validation procedures. Qualification documentation, including installation qualification and operational qualification, is standard practice in pharmaceutical research environments where IVRT and IVPT data are intended for regulatory submissions.<\/p>\n<h2>How ChemSPX supports IVRT and IVPT workflows in the Benelux<\/h2>\n<p>For laboratories in the Benelux region looking to set up or optimize IVRT and IVPT capabilities, ChemSPX provides specialist support across the full workflow. The team combines instrument expertise with application-level knowledge to help researchers select, install, and qualify the right setup for their specific formulation testing needs.<\/p>\n<ul>\n <li>Advice on instrument selection for static and automated Franz diffusion cell systems<\/li>\n <li>Support with IQ\/OQ installation and qualification in compliance with GMP and GLP requirements<\/li>\n <li>Application-level guidance for semisolid formulation testing workflows<\/li>\n <li>Ongoing service and calibration to keep instruments performing reliably over time<\/li>\n<\/ul>\n<p>Whether you are building a new topical formulation testing lab or expanding an existing analytical setup, the ChemSPX team is available to advise. <a href=\"https:\/\/chemspx.com\/nl\/contact\/\">Contact ChemSPX<\/a> to discuss your IVRT or IVPT requirements with a specialist.<\/p>\n<div class=\"wp-block-seoaic-faq-block\">\n    <h2 class=\"seoaic-faq-section-title\">Veelgestelde vragen<\/h2>\n            <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                How long does a typical IVRT or IVPT experiment take from setup to final data collection?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Most IVRT experiments run for 6 to 8 hours, with sampling points taken at regular intervals throughout. IVPT studies often run longer, sometimes up to 24 hours, because skin membranes can introduce a lag time before drug permeation becomes measurable. Planning for this lag phase is important when designing your sampling schedule, as collecting too few early time points can make it difficult to accurately calculate flux and permeation rates.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What are the most common mistakes made when setting up a Franz diffusion cell experiment?            <\/h3>\n            <p class=\"seoaic-answer\">\n                The most frequent issues are poor membrane mounting, inadequate sink conditions in the receptor fluid, and inconsistent temperature control. If the membrane is not sealed correctly against the cell joint, drug can bypass the membrane entirely and produce falsely elevated results. Equally, if the receptor fluid becomes saturated with drug before the end of the experiment, the concentration gradient collapses and permeation appears to slow down artificially. Using a validated receptor fluid composition and checking for leaks before each run are simple but critical steps.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                Can IVRT or IVPT be used for cosmetic formulations, or are they limited to pharmaceutical products?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Both tests are widely used for cosmetic formulations, particularly for products that claim skin penetration or active ingredient delivery, such as anti-aging creams, sunscreens, and topical actives like retinol or niacinamide. In the cosmetic industry, IVPT is especially useful for substantiating marketing claims about ingredient absorption without conducting invasive clinical studies. Regulatory requirements differ from pharmaceuticals, but the experimental methodology is essentially the same.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                How do you choose the right receptor fluid for IVRT and IVPT experiments?            <\/h3>\n            <p class=\"seoaic-answer\">\n                The receptor fluid must maintain sink conditions throughout the experiment, meaning the drug concentration in the receptor should remain well below its saturation solubility so that diffusion continues unimpeded. For most pharmaceutical IVRT studies, a phosphate-buffered saline solution at physiological pH is used. For lipophilic compounds that have poor aqueous solubility, a small percentage of organic solvent or a surfactant such as PEG 20 oleyl ether may be added. The receptor fluid composition should be validated to confirm that it does not artificially accelerate or retard drug transport.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What statistical considerations are important when interpreting IVRT or IVPT data for regulatory submissions?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Regulatory agencies typically require a minimum number of replicates per formulation, often six or more Franz diffusion cells per test group, to generate statistically meaningful flux and release rate data. The primary endpoint is usually the cumulative amount of drug permeated or released per unit area over time, from which a steady-state flux and lag time are calculated. For bioequivalence studies using IVRT, the 90% confidence interval for the ratio of release rates between test and reference products must fall within an acceptable range, similar in principle to pharmacokinetic bioequivalence criteria.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                Is it possible to run IVRT and IVPT in the same laboratory with the same Franz diffusion cell equipment?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Yes, the same Franz diffusion cell hardware can be used for both IVRT and IVPT by simply changing the membrane type between experiments. The transition requires thorough cleaning and validation of the cells between different membrane types to avoid cross-contamination. Laboratories running both test types routinely dedicate specific cell sets to each application or implement a rigorous cleaning and documentation protocol to maintain data integrity across both workflows.            <\/p>\n        <\/div>\n                <div class=\"seoaic-faq-item\">\n            <h3 class=\"seoaic-question\">\n                What should a laboratory consider before submitting IVRT or IVPT data as part of a regulatory dossier?            <\/h3>\n            <p class=\"seoaic-answer\">\n                Before including IVRT or IVPT data in a regulatory submission, laboratories should confirm that the method has been fully validated according to applicable guidelines, including specificity, linearity, accuracy, and precision of the analytical quantification method. The Franz diffusion cell equipment must be qualified under IQ\/OQ protocols, and all raw data should be traceable and audit-ready. It is also advisable to review the specific product guidance issued by the relevant regulatory agency, such as the FDA&#8217;s product-specific guidances for topical generics, as membrane type, receptor fluid, and study design requirements can vary by active ingredient and formulation category.            <\/p>\n        <\/div>\n        <\/div>","protected":false},"template":"","class_list":["post-8111","wiki_post","type-wiki_post","status-publish","has-post-thumbnail","hentry"],"acf":[],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v28.6 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>What is the difference between IVRT and IVPT for topical formulations? | ChemSPX<\/title>\n<meta name=\"description\" content=\"IVRT and IVPT serve different purposes in topical formulation testing. Discover what each measures, when regulators require it, and how both tests work together.\" \/>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/\" \/>\n<meta property=\"og:locale\" content=\"nl_NL\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"What is the difference between IVRT and IVPT for topical formulations? | ChemSPX\" \/>\n<meta property=\"og:description\" content=\"IVRT and IVPT serve different purposes in topical formulation testing. Discover what each measures, when regulators require it, and how both tests work together.\" \/>\n<meta property=\"og:url\" content=\"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/\" \/>\n<meta property=\"og:site_name\" content=\"ChemSPX\" \/>\n<meta property=\"og:image\" content=\"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/10\/franz-diffusion-cell-membrane-permeation-lab.webp\" \/>\n\t<meta property=\"og:image:width\" content=\"1024\" \/>\n\t<meta property=\"og:image:height\" content=\"768\" \/>\n\t<meta property=\"og:image:type\" content=\"image\/webp\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Geschatte leestijd\" \/>\n\t<meta name=\"twitter:data1\" content=\"10 minuten\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/\",\"url\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/\",\"name\":\"What is the difference between IVRT and IVPT for topical formulations? | ChemSPX\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/10\\\/franz-diffusion-cell-membrane-permeation-lab.webp\",\"datePublished\":\"2026-10-24T06:00:00+00:00\",\"description\":\"IVRT and IVPT serve different purposes in topical formulation testing. Discover what each measures, when regulators require it, and how both tests work together.\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/#breadcrumb\"},\"inLanguage\":\"nl-NL\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"nl-NL\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/#primaryimage\",\"url\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/10\\\/franz-diffusion-cell-membrane-permeation-lab.webp\",\"contentUrl\":\"https:\\\/\\\/chemspx.com\\\/wp-content\\\/uploads\\\/2026\\\/10\\\/franz-diffusion-cell-membrane-permeation-lab.webp\",\"width\":1024,\"height\":768,\"caption\":\"Franz diffusion cell apparatus with amber fluid in dual chambers, translucent membrane, sample vials, and pipette on laboratory bench.\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/wiki\\\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/chemspx.com\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Wiki post\",\"item\":\"https:\\\/\\\/chemspx.com\\\/Wiki\\\/\"},{\"@type\":\"ListItem\",\"position\":3,\"name\":\"What is the difference between IVRT and IVPT for topical formulations?\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/#website\",\"url\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/\",\"name\":\"ChemSPX\",\"description\":\"ChemSPX\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/chemspx.com\\\/nl\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"nl-NL\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"What is the difference between IVRT and IVPT for topical formulations? | ChemSPX","description":"IVRT and IVPT serve different purposes in topical formulation testing. Discover what each measures, when regulators require it, and how both tests work together.","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/","og_locale":"nl_NL","og_type":"article","og_title":"What is the difference between IVRT and IVPT for topical formulations? | ChemSPX","og_description":"IVRT and IVPT serve different purposes in topical formulation testing. Discover what each measures, when regulators require it, and how both tests work together.","og_url":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/","og_site_name":"ChemSPX","og_image":[{"width":1024,"height":768,"url":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/10\/franz-diffusion-cell-membrane-permeation-lab.webp","type":"image\/webp"}],"twitter_card":"summary_large_image","twitter_misc":{"Geschatte leestijd":"10 minuten"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"WebPage","@id":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/","url":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/","name":"What is the difference between IVRT and IVPT for topical formulations? | ChemSPX","isPartOf":{"@id":"https:\/\/chemspx.com\/nl\/#website"},"primaryImageOfPage":{"@id":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/#primaryimage"},"image":{"@id":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/#primaryimage"},"thumbnailUrl":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/10\/franz-diffusion-cell-membrane-permeation-lab.webp","datePublished":"2026-10-24T06:00:00+00:00","description":"IVRT and IVPT serve different purposes in topical formulation testing. Discover what each measures, when regulators require it, and how both tests work together.","breadcrumb":{"@id":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/#breadcrumb"},"inLanguage":"nl-NL","potentialAction":[{"@type":"ReadAction","target":["https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/"]}]},{"@type":"ImageObject","inLanguage":"nl-NL","@id":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/#primaryimage","url":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/10\/franz-diffusion-cell-membrane-permeation-lab.webp","contentUrl":"https:\/\/chemspx.com\/wp-content\/uploads\/2026\/10\/franz-diffusion-cell-membrane-permeation-lab.webp","width":1024,"height":768,"caption":"Franz diffusion cell apparatus with amber fluid in dual chambers, translucent membrane, sample vials, and pipette on laboratory bench."},{"@type":"BreadcrumbList","@id":"https:\/\/chemspx.com\/nl\/wiki\/what-is-the-difference-between-ivrt-and-ivpt-for-topical-formulations\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/chemspx.com\/"},{"@type":"ListItem","position":2,"name":"Wiki post","item":"https:\/\/chemspx.com\/Wiki\/"},{"@type":"ListItem","position":3,"name":"What is the difference between IVRT and IVPT for topical formulations?"}]},{"@type":"WebSite","@id":"https:\/\/chemspx.com\/nl\/#website","url":"https:\/\/chemspx.com\/nl\/","name":"ChemSPX","description":"ChemSPX","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/chemspx.com\/nl\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"nl-NL"}]}},"_links":{"self":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/wiki_post\/8111","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/wiki_post"}],"about":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/types\/wiki_post"}],"wp:featuredmedia":[{"embeddable":true,"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/media\/8156"}],"wp:attachment":[{"href":"https:\/\/chemspx.com\/nl\/wp-json\/wp\/v2\/media?parent=8111"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}