Choose your region:

Teledyne Labs

Vision AutoPlus with AutoFill

Seamless integration, intuitive programming and operation, flexibility to handle a wide range of applications
Choice of 6 or 8 sampling syringes (for 6 or 8 vessels)
Collect in test tubes for UV or vials for HPLC & UPLC
All-inert syringes, valves, tubing, and fluidics

Vision AutoPlus with AutoFill

Autosampler

The Vision® AutoPlus™ offers a choice of 6 or 8 sampling channels and can be configured with either DissoScan™ (1 bath) or Maximizer™ (up to 3 baths). Each unit comes standard with precision syringe pump mechanics, all-inert tubing and valves, and color touchscreen interface capable of storing up to 100 protocols.

Features of the AutoPlus autosampler:

  • Precision syringe pump autosampling system
  • All-inert syringes, valves, tubing, and fluidics
  • Choice of 6 or 8 sampling syringes for 6- or 8-vessel configurations
  • Volume accuracy and repeatability of ±1%
  • Seamless integration with G2 Classic 6™ and Elite 8™ systems
  • Automatic media replacement standard with Maximizer™
  • Auto Start functionality
  • Automatic addition of dilutions or stabilizers
  • Optional in-line filter block kit for simplified tubing and filter connections
  • Handles difficult formulations, surfactants, and buffers
  • Maintenance-friendly design without peristaltic pumps or check valves
  • Compact stacked footprint preserving bench space
  • Customizable end-of-test auto-wash routine
  • Full-color touchscreen with intuitive menus and programming
  • Storage of up to 100 protocols
  • Upload and download functionality via flash drive
  • 32-bit digital electronics and firmware
  • Printing support for serial and network (PostScript) printers
  • Sampling, maintenance, and calibration alarms
  • Compatible with non-Hanson dissolution testers
  • Security support for up to 50 users
  • 21 CFR Part 11 compliant security

Sample Collector

The Vision® AutoFill™ is a sample collector designed for seamless integration with the Vision AutoPlus autosampler, saving space while providing a full 360 degree view for visual monitoring of the collection process. It supports a variety of accessories, including collection racks with 13 x 100 or 16 x 100 mm test tubes for UV-Vis, 12 x 32 mm vials for HPLC, and sample trays with 12 x 32 mm vials for UPLC®, among others.

Features:

  • Precision fraction collector built for the Vision® platform
  • Seamless integration with AutoPlus™
  • Choice of easy install collection racks with rack detection sensors
  • Collect in test tubes for UV or vials for HPLC & UPLC®​
  • Compatible with 13 x 100 and 16 x 100 mm test tubes and 12 x 32 mm vials
  • New wash port with automatic end-of-test wash routine
  • Sample retrieval return to vessel accessory
  • CE / CSA / RoHS compliant, includes earth-friendling packing

Autosampling​ Configurations

Vision G2 dissolution test instruments are built for seamless integration, intuitive programming, and flexible operation, making them suitable for a wide range of applications while maintaining maximum throughput and productivity. Lab systems can be configured to handle anything from basic single-bath routines to extensive multi-bath testing.

The system also enables sample collection for off-line analysis in UV-Vis, HPLC, or UPLC® detectors. The product line includes the Vision G2 Classic 6, Vision G2 Elite 8, as well as the Vision AutoPlus autosampler and AutoFill collector.​

Collection Racks

AutoFill collection racks feature smart technology that alerts the AutoPlus autosampler if the wrong rack type is used for a method. In such cases, the AutoPlus displays a notification message, ensuring the correct rack is selected for testing. A variety of rack styles are available, including options for 16 x 100 mm test tubes, 13 x 100 mm test tubes, and 12 x 32 mm HPLC Vials.

Additional options include collection racks, vial trays, and needle block kits compatible with Agilent HPLC systems, as well as Waters Acquity UPLC systems. Test tubes and vials with pre-cut septum are also available.

Specifications 

AutoPlus

Weight (dry) 21.3 kg (47.0 lb)
Dimensions (H × W × D) 46.8 × 38.1 × 66.0 cm (18.4 × 15.0 × 26.0 in.)
Power Input: 24 VDC, 3.75 A, 90 W
External power supply required

AutoFill

General Weight (dry) 14.5 kg (32.0 lb)
Dimensions (H × W × D) 40.1 × 28.0 × 47.3 cm (15.8 × 11.0 × 18.6 in.)
Sampling Minimum sample interval 1 min (5 min typical)
Maximum sample interval 999 hours, 59 min (999:59)
Sample time points Max. 50
Maximum sample volume Up to 90 % of vial capacity
Collection Rack options 13 × 100 rack: 18 × 8 test tubes (144 total)
16 × 100 rack: 18 × 8 test tubes (144 total)
12 × 32 rack: 21 × 8 vials (168 total)
UPLC rack (12 × 32): 3 trays, 6 × 8 vials (144 total)
Septum type Pre‑cut type only
Sampler volume accuracy 0.5–5.0 mL ± 0.05 mL
5.1–10.0 mL ± 0.10 mL
10.1–15.0 mL ± 0.15 mL
Power (instrument) Input 24 VDC, 3.75 A, 90 W
Requirement External power supply only
External power supply Input 100–240 V~, 50–60 Hz, max. 1.5 A
Output 24 VDC, 3.75 A, 90 W

Questions about your application?

Our Xperts are happy to answer your questions and provide additional information about this instrument and its applications.

Frequently asked questions

The Vision® AutoPlus™ is an automated dissolution sampling system that upgrades the legacy Hanson AutoPlus™ platform. It provides precise, automated sampling with either 6 or 8 sampling channels and integrates directly with Vision G2 dissolution testers for efficient, high‑throughput workflows.
The Vision® AutoPlus™ automates sample withdrawal, collection, and media handling during dissolution testing. It replaces manual sampling, improving reproducibility, reducing operator workload, and enabling consistent sample timing across multiple vessels.
The Vision® AutoFill™ is a fraction collector designed to work with the AutoPlus autosampler. It collects samples into tubes or vials and allows full 360-degree visibility for monitoring the collection process.
The AutoFill supports multiple formats including 13 × 100 mm and 16 × 100 mm test tubes for UV‑Vis analysis, as well as 12 × 32 mm vials for HPLC and UPLC workflows.
The system uses rack detection sensors to confirm that the correct collection rack is installed. If the wrong rack is used, the system alerts the operator, reducing errors in sample collection.
The Vision® AutoPlus™ combines precise syringe-based sampling, flexible configuration, regulatory compliance, and seamless integration with dissolution testing systems. Together with the AutoFill collector, it enables efficient, automated, and high‑throughput dissolution workflows.

Downloads

Vision Autoplus Brochure

Related instruments

Vision G2 Elite 8

Dissolution tester suitable for both automated workflows and extended applications, as well as standard 8-sample manual testing
Speed range: 25 to 250 rpm
Temperature range: 25 to 55 °C

Vision G2 Classic 6

Compact and robust dissolution tester
Speed range: 25–250 rpm
Temperature range: 25–55 °C

Last viewed products